Gas |
|||||
GAS
Point of Use |
Device on Reactor
|
||||||||||
Gas |
In state |
Conver sion factor |
Out max |
working pressure in bar |
In max in bar |
Name of Device |
Differential pressure in bar |
Out max in sccm |
Out min in sccm |
Out state |
|
from |
to |
||||||||||
Air |
Gas |
8,0 |
6,0 |
Entrance chamber for flushing in Air |
Gas |
||||||
8,0 |
Pneumatic valves | ||||||||||
Nitrogen
|
Gas |
1,000 |
6,8
|
3,0 |
Entrance chamber for flushing in Nitrogen |
Gas |
|||||
6,8 |
Pressure for liquid precursor | ||||||||||
We use Flowmeter 1 for Nitrogen, even if Flowmeter 1 has been calibrated for Oxygen, because the conversion factors of Nitrogen and Oxygen are similar (and Nitrogen and Oxygen can be mixed) | Gas |
||||||||||
Oxygen |
Gas |
0,993 |
6,8 |
3,0 |
6,8 |
Flowmeter 1 (MKS 1179B) calibrated for Oxygen |
0,7 |
3,8 |
200 |
10 |
Gas |
Argon |
Gas |
1,390 |
6,8 |
3,0 |
6,8
|
Flowmeter 2 (MKS 1179B) calibrated for Argon |
0,7 |
3,8 |
200 |
10 |
Gas |
Helium |
Gas |
1,450 |
6,8 |
3,0 |
We use Flowmeter 2 for Helium, even if Flowmeter 2 has been calibrated for Argon, because the conversion factors of Helium and Argon are similar (and Argon and Helium can be mixed) | Gas |
|||||
Hydrogen |
Gas |
1,010 |
6,8 |
3,0 |
6,8 |
Flowmeter 3 (MKS 1179B) calibrated for Hydrogen |
0,7 |
3,8 |
200 |
10 |
Gas |
Shuttle |
Device on Reactor |
||||||||
Precursor |
In state |
Out max in relative bar |
In max in relative bar |
Name of Device |
Differential pressure in bar |
Out max in l/hr |
Out min in l/hr |
Out state |
|
from |
to |
||||||||
TiCl4 |
Liquid | 6,8 |
6,8 |
Flowmeter (Quantim QMBC) for liquid precursor | 0,7 |
10,3 |
0,30 |
0,01 |
Gas |